Replacing the APCI probe sample capillary
5-77
3. Remove the end cover and gasket.
4. Unscrew and remove the nebulizer adjuster knob.
5. Remove the PEEK union/UNF coupling assembly and capillary from the
probe.
6. Use the 7-mm wrench to loosen the locknut.
Nebulizer adjuster knob
End cover
Gasket
PEEK union/UNF coupling assembly
Capillary
Locknut