Replacing the APCI probe sample capillary
5-81
18. Use the combined 2.5-mm Allen wrench and cone extraction tool to fit
and tighten the 3 end-cover securing screws.
19. Fit the probe heater (see
).
20. Fit the probe to the instrument (see
).
21. In the Instrument Console, click API
to start the probe and
desolvation gas flows.
Caution:
• When handling the probe heater, take great care to grip the
heater so as not to damage its electrical wiring.
• Take great care not to damage the probe heater’s electrical
connections, capillary sleeve, or capillary when fitting the
heater over the capillary sleeve.