Cleaning the ion block assembly
5-55
6. Inspect each component for persisting contamination.
Requirement:
If contamination is present, repeat the cleaning
procedure. If contamination persists, dispose of the component, and
obtain a new one before reassembling the ion block assembly.
Assembling the source ion block assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
Needle-nose pliers
•
Isopropyl alcohol in small container
•
Lint-free cloth
To assemble the ion block assembly:
1. Fit the extraction cone seal to the ion block.
2. Fit the extraction cone handle insulator to the extraction cone handle.
3. Fit the extraction cone to the ion block.
4. Fit the 2 PEEK extraction cone retainer clips to the ion block.
Warning:
The components can be contaminated with
biohazardous and/or toxic materials. Dispose of them
according to local environmental regulations.
Caution:
• To avoid recontaminating the ion block assembly, wear clean
chemical-resistant, powder-free gloves during this procedure.
• The extraction cone is fragile. Never place it on its tip; always place it
on its flanged base.
Caution:
Take great care not to damage the extraction cone
aperture when fitting the extraction cone to the ion block.
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...