Installing the small-bore capillary option
3-11
19. Fit a new, metal gasket to the probe tip.
20. Fit the probe tip over the capillary, and screw the tip onto the probe
assembly.
21. Use the 10-mm wrench to tighten the probe tip.
Caution:
To avoid gas leakage, fully tighten the probe tip.
Metal gasket
Probe tip
10-mm wrench