3-14
Configuring the LockSpray Source
To install the APCI probe:
1. Prepare the instrument for working on the source (see
2. With the probe label facing you, carefully slide the APCI probe into the
hole in the probe adjuster assembly, ensuring that the probe location
dowel aligns with the location hole in the probe adjuster assembly.
Warning:
The LC system connections, APCI probe, and source can
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this procedure.
TP03129
APCI probe location dowel
Location hole in the probe
adjuster assembly
Probe label