Cleaning the ion block assembly
5-49
11. Use the needle-nose pliers to grasp the PEEK terminal block and
partially lift it out of the ion block.
12. Holding the PEEK terminal block, gently use the needle-nose pliers to
gently grasp the heat-shrink tubing on the heater cartridge assembly,
and carefully slide it and the PEEK terminal block out of the ion block.
Caution:
To avoid damaging the heater cartridge assembly wires,
do not bend or twist them when removing the assembly from the
ion block.
PEEK terminal
block securing
screw
Heat-shrink tubing
Heater cartridge assembly
PEEK terminal block