![ThermoFisher Scientific thermo scientific imPULSE S.U.M. User Manual Download Page 101](http://html1.mh-extra.com/html/thermofisher-scientific/thermo-scientific-impulse-s-u-m/thermo-scientific-impulse-s-u-m_user-manual_1105607101.webp)
Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
96
Chapter 3
|
Touchscreen Console setup
2. To set the High/Low alarm, move the corresponding slider to the
On
position and enter the value that will activate the High/Low
alarm.
3. Touch
Enter
to confirm and return to the previous screen.
4. After you have entered the High and Low alarm values, touch
Done
.
5. After a moment, a confirmation screen will appear. Touch
Done
to return to the Auxiliary Input Module Configuration screen—the
values in the High and Low field will reflect your changes.
High High and Low Low alarms
Choose to enable or disable High High and/or Low Low alarms.
These alarms provide a visual and audible alert, and will pause or stop
specified pumps based on the set Interlock equations.
1. If you would like to set High High/Low Low alarms for the Auxiliary
Input module, touch the
High High and Low Low alarms
field.
2. On the screen that appears, there are separate sections for both
High High and Low Low alarms, each with respective Interlock
and Hysteresis fields (Figure 3.76). To set either alarm, move the
corresponding slider to the
On
position.
Figure 3.75. Aux Input High and Low alarms screen.