Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
98
Chapter 3
|
Touchscreen Console setup
Figure 3.77. Aux Input High and Low alarms screen.
Note:
Enabling a pump allows it to adjust the conductivity level by
delivering a bolus after the alarm has been triggered.
2. On the screen that appears (Figure 3.77), you will see the interlock
and hysteresis parameters you entered for the High High and/or
Low Low alarm(s). Below that, the pumps that have been set up
with the system are shown. Touch a
pump
option.
3. On the screen that appears, move the corresponding slider to the
On
position to turn on the High High and/or Low Low alarm(s) for
the selected pump.
4. Repeat this step to assign alarms to any other pumps that have
been set up. When you are finished, touch the
Done
button.
5. After setting all desired alarm designations touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
to
return to the Auxiliary Configuration screen—the values in the Alarm
designation field will reflect your changes.