Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
163
Table 8.10. Touchscreen Console measurement options and specifications.
Load cell
Temperature
pH
Conductivity
BPC and in-line
liquid pressure
Hz accuracy
Accuracy after
calibration
± 0.5% of full scale
0.2°C
± 0.05 pH
± 5%
± 3.5% of full
scale (30 psi)
0.02 Hz or
5% of setpoint
(whichever is
greater)
Calibration
1 to 3 points and zero/tare
function
1 to 3 points
1 to 3 points
1 to 3 points 1 point; zero/tare
1 point offset
Resolution
0.1 kg
0.01°C
0.01 pH
1 µS/cm
0.01 psi
0.5 Hz
±
0.03 Hz
1.0 Hz
±
0.05 Hz
1.5 Hz
±
0.08 Hz
2.0 Hz
±
0.10 Hz
Sensor range
30 L: 440 kg ± 1.4 kg
100 L: 880 kg ± 3.1 kg
250 L: 880 kg ± 3.1 kg
500 L: 2,200 kg ± 7.7 kg
1,000 L: 4,400 kg ± 15.4 kg
2,000 L: 3,300 kg ± 11.5 kg
3,000 L: 6,600 kg ± 23.1 kg
5,000 L: 13,200 kg ± 46.2 kg
0–200°C
0–14
20–20,000
µS/cm
0–30 psi
0–2 Hz
Measurement
units
kg
Degrees
Celsius
pH
µS/cm
psi
Hz
Probe type
Mettler Toledo MTB load cells
QTY 3 or 4
RTD
Electrochemical
with 225 mm
S8 connector
Two-pole
conductivity
sensor
Single-use
sensor, part
of BPC and/
or fluid transfer
assembly design
N/A
Chapter 8
|
Specifications and parts information
8.4 BPC specifications
All imPULSE S.U.M.s are constructed with ASI film and silicone tubing.
Figure 8.18 shows a wireframe representation of a standard imPULSE
BPC.