Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
39
Chapter 3
|
Touchscreen Console setup
filter). The Liquid Pressure module is used to monitor the liquid
pressure level during harvest to ensure that the filter is not clogged.
The use of pumps and/or a Harvest pinch valve is optional with
this module: the pinch valve will close and the pump will stop if the
liquid pressure reads above a specified setpoint.
• Auxiliary Input
—AuxIn2 can be used to monitor sensors and
transmitters that use a 4-20 mA signal, such as external scales.
Note:
AuxIn1 is reserved for the laser distance sensor, used to
home the carriage to either the top or bottom position. As such, it is
grayed out and cannot be used in the same manner.
3.3 Required external components for modules
The flow chart provided below (Figure 3.9) illustrates the required
external components for various modules.
pH Sensor
Conductivity
Sensor
Air Pressure
Sensor
Air Source
Load Cells
Pinch Clamps
Temperature
Liquid Pressure
Sensor
Pump(s)
Temperature
Sensor
Agitation
Auxiliary
Inputs
Timer
Harvest
pH
Conductivity
BPC Pressure
Mass
Fill
Liquid Pressure
PCD Pressure
Auxiliary
Outputs
Pumps
Module
External Component
Requires
Figure 3.9. Flow chart depicting required external components for each module.