Chapter 3
|
Touchscreen Console setup
1. Touch the
Scaling
field.
2. On the screen that appears, enter the scaling you would like to use
between the mA signal received by the Touchscreen Console, and
the specified units (Figure 3.74).
Figure 3.74. Aux Input Scaling screen.
3. When you are finished, touch
Done
.
4. After a moment a confirmation screen will appear. Touch
Done
again to return to the Auxiliary Input Module Configuration screen.
Optional fields
High and Low alarms
Enter the pH values that will trigger High and/or Low pH alarms. These
alarms provide a visual alert on the display when conditions are outside
of specified operating parameters.
1. If you would like to set High/Low alarms for the Auxiliary Input
module, touch the
High and Low alarms
field. On the screen that
appears (Figure 3.75), there will be separate sections for both High
and Low alarms.