imPULSE S.U.M. with Touchscreen Console User’s Guide
|
56
Chapter 3
|
Touchscreen Console setup
Figure 3.26. BPC Pressure High and Low Alarms screen.
Optional fields
High and Low alarms
Choose to enable or disable High and/or Low alarms. These alarms
provide a visual alert on the display when conditions are outside of
specified operating parameters.
1. If you would like to set High/Low alarms for air pressure, touch the
High and Low alarms
field. The screen that appears has separate
sections for both High and Low alarms (Figure 3.26).
2. To set either alarm, move the corresponding slider to the
On
position.
3. Touch the High or Low alarm field and enter the pressure value (in
either mbar or PSI) that will trigger the High or Low alarm.
4. Touch
Enter
to confirm and return to the previous screen.
5. After you have entered the High and Low alarm values, touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
to
return to the Configuration screen—the values in the High and Low
alarms field will reflect your changes.
High High and Low Low alarms
Choose to enable or disable High High and/or Low Low alarms.
These alarms provide a visual and audible alert, and will pause or stop
specified pumps based on the set Interlock equations.