Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
79
Chapter 3
|
Touchscreen Console setup
High and Low alarms
Enter the pH values that will trigger High and/or Low pH alarms. These
alarms provide a visual alert on the display when conditions are outside
of specified operating parameters.
1. If you would like to set High/Low alarms for pH, touch the
High and
Low alarms
field. On the screen that appears, there are separate
sections for both High and Low alarms (Figure 3.57).
Figure 3.57. pH High and Low Alarms screen.
2. To set either alarm, move the corresponding slider to the
On
position.
3. Touch the High or Low alarm field and enter the pH values that will
trigger the alarm.
4. Touch
Enter
to confirm and return to the previous screen.
5. After you have entered the High and Low alarm values, touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
to
return to the pH Module Configuration screen—the values in the
High and Low alarms field will reflect your changes.
High High and Low Low alarms
Choose to enable or disable High High and/or Low Low alarms.
These alarms provide a visual and audible alert, and will pause or stop
specified pumps based on the set Interlock equations.