Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
35
Chapter 3
|
Touchscreen Console setup
7. Ensure that the date and time displayed in the bottom right corner
of the idle screen are correct. If either value is incorrect, touch
the
Settings
icon in the top right corner and touch
Instrument
Settings
to adjust the values. You can also change the Instrument
name that appears at the top of the Home screen.
Note:
To make changes to these settings after configuration,
touch the Settings icon at the top right and select
Mixer & module
configurations.
Then select
Actions
and
Edit
modules. If an
Administrator account has been created, changes can only be made
when signed in as the Administrator.
3.1.3 Creating and using an Administrator profile
An Administrator profile allows an administrator to limit other
users’ accessibility of certain Touchscreen Console settings and
functions. Table 3.2 shows the available functions for all users and for
administrators only.
Table 3.2. Accessibility of settings/functions for all users and administrators only.
Setting or function
All users
Administrators only
pH settings
Calibration, Temperature compensation,
Select active sensor, Reset calibration,
Module configuration, and Sensors available
High and Low alarms, High High and Low Low alarms, Alarm
designation (pumps), and Diagnostics
System settings
Export sensor logs
About instrument, Maintenance and services, Instrument
settings, Alarms and interlocks, and End User License
Agreement (EULA)
BPC pressure settings
Units, Calibration
High and Low alarms, High High and Low Low alarms, Alarm
designation (pumps), and Module Configuration
Agitation settings
N/A
High and Low alarms, High High and Low Low alarms,
Setpoint limits, Calibration, and Module configuration
Temperature settings
Select active sensor, Calibration
High and Low alarms, High High and Low Low alarms, TCU
setpoint limits, and Module configuration
Timer settings
Notes, Reset
N/A
Mass settings
Fluid density, Tare/zero
High and Low alarms, High High and Low Low alarms,
Alarm designation (pumps), Alarm designation (agitation),
Alarm designation (pinch valves), Calibration, and Module
configuration
Pump settings
Units and fluid density, Pump calibration,
Add bolus, and Prime pump
Diagnostics, RPM scaling, RPM limits, and Module
configuration
Liquid pressure settings Units, Display options, and Calibration
High and Low alarms, High High and Low Low alarms, Alarm
designation (pumps), Alarm designation (agitation), Alarm
designation (pinch valves), and Module configuration
PCD Pressure
Types & Units and Calibration
High and Low alarms, High High and Low Low alarms, Alarm
designation (pumps), and Module configuration