Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
43
Chapter 3
|
Touchscreen Console setup
3. After entering the setpoint limits, touch
Done
.
4. A confirmation screen will appear. Touch
Done
to return to the
Configuration screen—the values in the
Set point limits
field will
reflect your changes.
Figure 3.13. Agitation High and Low alarms screen.
Optional fields
High and Low alarms
Choose to enable or disable High and/or Low alarms. These alarms
provide a visual alert on the display when conditions are outside of
specified operating parameters.
1. If you would like to set High/Low alarms for agitation, touch the
High and Low alarms
field. The screen that appears has separate
sections for High and Low alarms (Figure 3.13).
2. To set either alarm, move the corresponding slider to the
On
position.
3. Touch the
High
or
Low alarm
field and enter the value (in Hz) that
will trigger the High or Low alarm.
4. Touch
Enter
to confirm and return to the previous screen.
5. After you finish setting the High and Low alarm values, touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
to
return to the Configuration screen—the values in the
High and Low
alarms
field will reflect your changes.