Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
53
Chapter 3
|
Touchscreen Console setup
Figure 3.23. BPC Pressure High High and Low Low Alarms
screen.
3. Touch the
Interlock
field of an enabled alarm to enter the Interlock
equation screen.
4. Fill out the Interlock equation by entering the
Control setpoint
(in
either PSI or mbar) and selecting the
Action taken
(either
Pause
or
Stop
).
– If
Pause
is set as the action taken, you must fill out the Hysteresis
equation by entering the
Control setpoint
(in either PSI or mbar)
and the
Time to hold
before operations can resume (in seconds).
– If
Stop
is set as the action taken, operations must be resumed
manually.
5. Touch
Done
to return to the previous screen.
6. Once you finish setting the High High and/or Low Low alarms,
touch
Done
.
7. The system will confirm the settings. Touch
Done
to return to the
PCD Module Configuration screen—the values in the
High High
and Low Low
field will reflect your changes.
Alarm designation (pumps)
Use this field to specify which pump(s) are affected by the High High
and/or Low Low alarms.
1. If you would like to designate alarms for pumps that have been
set up, touch the
Alarm designation
(pumps)
field on the PCD
Module Configuration screen.