Chapter 4
|
Calibration procedures
4.6 Calibrating PCD pressure
If a PCD is present, use the following steps to calibrate PCD pressure.
1. Touch the top of the
PCD Pressure
module on the Touchscreen
Console Home screen to open the PCD Pressure Settings screen.
Touch
Calibration
.
2. A screen will appear prompting you to select a calibration method:
zero/tare, enter a 1-point offset value, or clear offset. Select
Enter a
1-point offset value
.
3. Connect a calibrated pressure gauge to a lineset and take a
reading in psi.
4. The following screen will prompt you to enter the BPC sensor
value in psi. Enter the reading from the pressure gage and touch
Calibrate
.
5. After the BPC Pressure Calibration Complete screen appears, the
calibration is complete. Touch
Done
in the lower right corner of the
screen to return to the BPC Pressure Settings screen.
4.7 Calibrating BPC air pressure
If a PCD is NOT present use the following steps to calibrate BPC air
pressure.
1. Touch the top of the
BPC Pressure
module on the Touchscreen
Console Home screen to open the BPC Pressure Settings screen.
Touch
Calibration
.
2. A screen will appear prompting the user to select a calibration
method: zero/tare, enter a 1-point offset value, or clear offset.
Select
Enter a 1-point offset value
.
3. Connect a calibrated pressure gauge to a lineset and take a
reading in psi.
4. The following screen will prompt you to enter the BPC sensor
value in psi. Enter the reading from the pressure gauge and touch
Calibrate
.
5. After the “BPC Pressure Calibration Complete” screen appears, the
calibration is complete. Touch
Done
in the lower right corner of the
screen to return to the “BPC Pressure Settings” screen.