Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
148
Chapter 8
|
Specifications and parts information
Table 8.2. 100 L imPULSE S.U.M. specifications.
Specification
Gen
er
al
Unit dimensions (D x W x H)
104.77 x 121.12 x 160.83 cm (41.25 x 47.68 x 63.30 in.)
Nominal working volume
100 L
Maximum working volume
107 L
Volume at probe belt
7.25
Minimum startup working volume
10%
Minimum drain working volume
0 L
Dry unit weight
303.4 kg (668.9 lbs)
Wet unit weight (nominal working volume)
410.18 kg (904.29 lbs)
Maximum noise level
< 78 dB at 1 m
Motor load max
1.32 A at 230 V (0.27 hp)
Electrical power supply requirement (U.S. systems)
AC motor module:
Powered by the TSC*
TSC:
10 A, 100–120/208–240 VAC, 50–60 Hz
Maximum temperature in vessel
40°C
In
fla
tio
n/
co
ve
r g
as u
tili
ty
re
qu
irem
en
ts
Maximum inflation operation air pressure
PCD:
80 psi
TSC:
95 psi
Temperature
Ambient
PCD inlet air connection
3/8 in. OD, push-to-connect tube
TSC inlet air connection
5/16 in. OD, push-to-connect tube
Process air pressure
30 psig**
Instrument air pressure
Regulate incoming air according to instrument specifications
Filtration
Disposable filtration integrated with optional tube sets
W
at
er u
tili
ty
re
qu
irem
en
ts
Jacket Flow rate
10–45 L/min.
Jacket Maximum temperature
50°C***
Jacket Minimum temperature
0°C
†
Connection inlet
3/4 in. NPT Quick Connect—SH6-63Y Male
Connection outlet
3/4 in. NPT Quick Connect—SH6-62Y Female
Ag
itat
io
n
Mixing disc diameter
21.6 cm (8.5 in.)
Mixer stroke
10.2 cm (4 in.)
Vessel turnover
7.44 L/sec. (13.5 total sec.)
Motor speed
0.5–2 Hz
*AC motor module nameplate voltage and TSC input voltage need to match.
**For manual inflation.
***The vessel jacket can be operated at 50°C until the BPC reaches 40°C, after which, the vessel jacket temperature must be lowered.
†
The vessel jacket can be operated at 0°C until the BPC approaches 5°C, after which, the vessel jacket temperature must be raised to match.