![ThermoFisher Scientific thermo scientific imPULSE S.U.M. User Manual Download Page 163](http://html1.mh-extra.com/html/thermofisher-scientific/thermo-scientific-impulse-s-u-m/thermo-scientific-impulse-s-u-m_user-manual_1105607163.webp)
Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
158
Chapter 8
|
Specifications and parts information
Table 8.7. 3,000 L imPULSE S.U.M. specifications.
Specification
Gen
er
al
Unit dimensions (D x W x H)
207.36 x 180.47 x 282.85 cm (81.64 x 71.05 x 111.36 in.)
Dimensions with optional BPC hoist (D x W x H)
207.36 x 208.10 x 413.39 cm (81.64 x 81.92 x 162.75 in.)
Nominal working volume
3,000 L
Maximum working volume
3,155 L
Volume at probe belt
242.64 L
Minimum startup working volume
10%
Minimum drain working volume
0 L
Dry unit weight
1,730 kg (3814 lb.)
Wet unit weight (nominal working volume)
4,534.89 kg (9,997.75 lb.)
Maximum noise level
< 78 dB at 1 m
Motor load max
18.5 A at 460 V (15 hp)
Electrical power supply requirement (U.S. systems)
AC motor module:
30 A, 400–480 VAC, 50–60 Hz
TSC:
15 A, 100–120/208–240 VAC, 50–60 Hz
BPC hoist:
12A, 100-120 VAC, 50-60 Hz; or 12A,
208-240 VAC, 50-60 Hz*
Maximum temperature in vessel
40°C
In
fla
tio
n/
co
ve
r g
as u
tili
ty
re
qu
irem
en
ts
Maximum inflation operation air pressure
PCD:
80 psi
TSC:
95 psi
Temperature
Ambient
PCD inlet air connection
3/8 in. OD, push-to-connect tube
TSC inlet air connection
5/16 in. OD, push-to-connect tube
Process air pressure
30 psig**
Instrument air pressure
Regulate incoming air according to instrument specifications
Filtration
Disposable filtration integrated with optional tube sets
W
at
er u
tili
ty
re
qu
irem
en
ts
Jacket Flow rate
10–45 L/min.
Jacket Maximum temperature
50°C***
Jacket Minimum temperature
0°C
†
Connection inlet
3/4 in. NPT Quick Connect—SH6-63Y Male
Connection outlet
3/4 in. NPT Quick Connect—SH6-62Y Female
Ag
itat
io
n
Mixing disc diameter
68.6 cm (27 in.)
Mixer stroke
20.3 cm (8 in.)
Vessel turnover
150.12 L/sec. (20 total sec.)
Motor speed
0–2 Hz
*Depending on specified hoist input voltage.
**For manual inflation.
***The vessel jacket can be operated at 50°C until the BPC reaches 40°C, after which, the vessel jacket temperature must be lowered.
†
The vessel jacket can be operated at 0°C until the BPC approaches 5°C, after which, the vessel jacket temperature must be raised to match.