Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
54
Chapter 3
|
Touchscreen Console setup
Note:
Enabling a pump allows it to adjust the air pressure after the
alarm has been triggered.
2. On the screen that appears (Figure 3.24), you will see the
interlock and hysteresis parameters that you entered for the High
High and/or Low Low alarm(s). Below that, the pumps that have
been set up with the system are shown. Touch a
pump
option.
3. On the screen that appears, move the corresponding slider to the
On
position for the High High and/or Low Low alarms to turn on the
alarm(s) for the selected pump.
4. When you are finished, touch
Done
to return to the Alarm
Designation screen.
5. After setting all desired alarm designations touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
to
return to the PCD Module Configuration screen—the values in the
Alarm designation field will reflect your changes.
Calibration
Set up a new PCD Pressure calibration using one-point offset.
1. If you would like to set up a PCD Pressure calibration, touch the
Calibration
field.
2. Follow the calibration instructions that are detailed in Chapter 4—
Calibration Procedures.
3. Touch
Done
to return to the Home screen.
Figure 3.24. BPC Alarm Designation (Pumps) screen.