![ThermoFisher Scientific thermo scientific imPULSE S.U.M. User Manual Download Page 117](http://html1.mh-extra.com/html/thermofisher-scientific/thermo-scientific-impulse-s-u-m/thermo-scientific-impulse-s-u-m_user-manual_1105607117.webp)
Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
112
Chapter 5
|
BPC loading instructions
Figure 5.12. Using two
hands to lift the BPC.
Figure 5.13. Shaft
facing down.
Figure 5.14. Placing the
BPC into the vessel.
Loading with the optional BPC hoist system (vessels 500 L or
larger)
1. Place the BPC with the BPC cradle below it on a rounded edge
table next to the vessel where the hoist lift can swing out over
it. This can be either toward the front or the back of the vessel
(Figure 5.16).
2. Press the down button on the hoist lift controller to lower the lift.
Lower the lift only as far as you need to attach it to the BPC cradle.
Do not lower past the floor creating slack in the hoist strap
as it can cause binding in the reel
.
3. Raise the BPC until it is over the top of the vessel and then swing
the arm so that the BPC is over the center of the vessel.
Leave
approximately 76.2–101.6 mm (3–4 in.) of the hoist strap
from the motor.
Figure 5.15. Use the lever shown
to swing the hoist lift arm.
Figure 5.16. Hoist lift arm in center, forward, and
back positions.