Specifications and parts information
Table 8.6. 2,000 L imPULSE S.U.M. specifications.
Specification
Gen
er
al
Unit dimensions (D x W x H)
189.58 x 162.42 x 259.92 cm (74.64 x 63.94 x 102.33 in.)
Height with optional BPC hoist (D x W x H)
189.58 x 190.05 x 384.43 cm (74.64 x 74.82 x 151.35 in.)
Nominal working volume
2,000 L
Maximum working volume
2,038 L
Volume at probe belt
192.40 L
Minimum startup working volume
10%
Minimum drain working volume
0 L
Dry unit weight
1,049 kg (2,411.9 lb.)
Wet unit weight (nominal working volume)
3,295.24 (7,264.72 lb.)
Maximum noise level
< 78 dB at 1 m
Motor load max
18.5 A at 460 V (15 hp)
Electrical power supply requirement (U.S. systems)
AC motor module:
30 A, 400–480 VAC, 50–60 Hz
TSC:
15 A, 100–120/208–240 VAC, 50–60 Hz
BPC hoist:
12A, 100-120 VAC, 50-60 Hz; or 12A,
208-240 VAC, 50-60 Hz*
Maximum temperature in vessel
40°C
In
fla
tio
n/
co
ve
r g
as u
tili
ty
re
qu
irem
en
ts
Maximum inflation operation air pressure
PCD:
80 psi
TSC:
95 psi
Temperature
Ambient
PCD inlet air connection
3/8 in. OD, push-to-connect tube
TSC inlet air connection
5/16 in. OD, push-to-connect tube
Process air pressure
30 psig**
Instrument air pressure
Regulate incoming air according to instrument specifications
Filtration
Disposable filtration integrated with optional tube sets
W
at
er u
tili
ty
re
qu
irem
en
ts
Jacket Flow rate
10–45 L/min.
Jacket Maximum temperature
50°C***
Jacket Minimum temperature
0°C
†
Connection inlet
3/4 in. NPT Quick Connect—SH6-63Y Male
Connection outlet
3/4 in. NPT Quick Connect—SH6-62Y Female
Ag
itat
io
n
Mixing disc diameter
61 cm (24 in.)
Mixer stroke
20.3 cm (8 in.)
Vessel turnover
118.61 L/sec. (16.6 total sec.)
Motor speed
0–2 Hz
*Depending on specified hoist input voltage.
**For manual inflation.
***The vessel jacket can be operated at 50°C until the BPC reaches 40°C, after which, the vessel jacket temperature must be lowered.
†
The vessel jacket can be operated at 0°C until the BPC approaches 5°C, after which, the vessel jacket temperature must be raised to match.