imPULSE S.U.M. with Touchscreen Console User’s Guide
|
59
Chapter 3
|
Touchscreen Console setup
3.4.7 Configuring the Pumps module
Use the following information and steps to configure the Pumps
module. To begin, locate the Pump module on the Touchscreen
Console Home screen, and touch
Configure
. The Pump Module
Configuration screen will appear (Figure 3.29).
Note:
You can touch
Done
at the bottom of the Configuration screen
at any time to save your current settings and so you can resume later.
Figure 3.29. Pump Configuration screen.
Required fields
RPM scaling
Convert the output signal for the pump from milliamps (mA) to RPM.
1. To change the scaling, touch the
RPM scaling
field.
Calibration
Set up a new BPC Pressure calibration using one-point offset.
1. If you would like to set up a BPC Pressure calibration, touch the
Calibration
field.
2. Follow the calibration instructions that are detailed in
Chapter 4—Calibration Procedures.
3. Touch
Done
to return to the Home screen.