Chapter 3
|
Touchscreen Console setup
Calibration
Calibration for the temperature sensors can be found in the
Temperature Module settings screen. Set calibration using one-, two-,
or three-point offset.
1. Touch the top of the Temperature Module to open the Temperature
Settings screen.
2. Follow the calibration instructions that are detailed in Section 4.9
3. If you are using a TCU, see the following section for TCU setup and
calibration parameters.
TCU setup and calibration parameters
The following graphics illustrate the setup and calibration parameters
for Lauda Integral T series, Lauda Varicool series, and Neslab TF series
TCUs.
Note:
Teal-colored boxes represent items from the Settings
menu, and gray-colored boxes represent the required parameter
inputs.
QUICK START/
SETTINGS
ANALOG
COMM
SERIAL COMM
SERIAL TYPE
REMOTE
SENSOR
RS 485
BAUD: 9600
PARITY: NON
DATA BITS: 8
STOP BITS: 1
ADDRESS: 1
Figure 3.47. Setup parameters for Neslab TF series TCU.
MENU
SETUP
CONTROL
INTERFACES
SERIAL
INTERFACES
MODE
RS 485
ADDRESS
WAIT A FEW
SECONDS TO
LOAD
SETTINGS
MENU
CONTROL
VARIABLE
EXTERN
PT100
RS485
1
REQUIRED
PARAMETER
INPUT
Figure 3.48. Setup parameters for Lauda Varicool series TCU.