Thermo Scientific
imPULSE S.U.M. with Touchscreen Console User’s Guide
|
81
Chapter 3
|
Touchscreen Console setup
1. If you would like to designate alarms for specific pumps that have
been set up for the S.U.M., touch the
Alarm designation
field on
the pH Configuration screen.
Note:
Enabling a pump allows it to adjust the pH level by delivering
a bolus after the alarm has been triggered.
2. On the screen that appears (Figure 3.59), you will see the interlock
and hysteresis parameters you entered for the High High and/or
Low Low alarm(s). Below that, the pumps that have been set up
with the system are shown. Touch a
pump
option.
Figure 3.59. pH Pump Designation screen.
3. On the screen that appears, move the corresponding slider to the
On
position for the High High and/or Low Low alarms to turn on the
alarm(s) for the selected pump.
4. Repeat this step to assign alarms to any other pumps that have
been set up.
5. When you are finished, touch
Done
.
6. After a moment, a confirmation screen will appear. Touch
Done
again to return to the Mass Module Configuration screen.
Calibration
Set up a new pH calibration using either direct entry, 2-point, or 3-point
calibration.
1. If you would like to calibrate the pH, touch the
Calibration
field.