Remote Commands
R&S
®
FPL1000
673
User Manual 1178.3370.02 ─ 09
MAX
The maximum of the relative
and
the absolute level is used as
RANGe<ri>:LIMit<li>:RELative:STOP
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STOP:ABS
commands to define these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative stop value. Use the
RANGe<ri>:LIMit<li>:RELative:STOP
this value.
*RST:
OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP:
FUNCtion MAX
Chapter 11.6.7.10, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:STATe
<State>
This command selects the limit check mode for
all
SEM ranges (<range> is irrelevant).
Suffix:
<sb>
.
1 to 3
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<State>
ABSolute | RELative | AND | OR
ABSolute
Checks only the absolute limits defined.
RELative
Checks only the relative limits. Relative limits are defined as rel-
ative to the measured power in the reference range.
AND
Combines the absolute and relative limit. The limit check fails
when both limits are violated.
OR
Combines the absolute and relative limit. The limit check fails
when one of the limits is violated.
*RST:
RELative
Manual operation:
See
Configuring and Performing Measurements