General Instrument Setup
R&S
®
FPL1000
525
User Manual 1178.3370.02 ─ 09
surements. The minimum frequency of a data point is 0 Hz, the maximum fre-
quency 200 GHz.
●
The value range for the transducer factor is ±200 dB.
●
Gain has to be entered as a negative value, and attenuation as a positive value.
Storing transducer factors
Transducer factors can also be stored with the configuration settings so they can be
recalled for other measurements at a later time. Note, however, that any changes
made to the transducer factors
after
storing the configuration file cannot be restored
and will be overwritten by the stored values when the configuration file is recalled.
Always remember to store the settings again after changing the transducer factors.
(See
Chapter 8.2, "Storing and Recalling Instrument Settings and Measurement Data"
Recalling transducer factors stored with measurement settings
After recalling measurement settings, the transducer factors applied to the measure-
ment may be different to those displayed in the "Transducer" dialog box; see
and recalling transducer and limit line settings"
9.7.2
Transducer Settings
Access
: [Setup] > "Transducer"
Up to 8 transducer lines can be activated simultaneously in the R&S
FPL1000. Many
more can be stored on the instrument.
Stored transducer settings
When storing and recalling transducer settings, consider the information provided in
"Saving and recalling transducer and limit line settings"
...................................................................................... 525
............................................................................................... 528
9.7.2.1
Transducer Management
Access
: [Setup] > "Transducer"
The settings required to manage all transducer lines on the instrument are described
here.
Transducers