The Spectrum Application (RF Measurements)
R&S
®
FPL1000
291
User Manual 1178.3370.02 ─ 09
Or
: [Freq] > "Generator Config"
The independent CW signal is available in all R&S
FPL1000 applications if the optional
Internal Generator R&S
FPL1-B9 is installed.
For details, see
Chapter 7.3.3.1, "Basics on the Internal Generator"
Usage
Determines the type of signal provided by the internal generator:
●
on page 283: constant frequency and power level
●
on page 284: constant power level, frequency follows ana-
lyzer frequency (only available for frequency sweeps in the Spectrum application)
●
on page 284: constant frequency, linearly increasing power level
(only available in the Spectrum application)
Remote command:
SOURce<si>:FREQuency:COUPling[:STATe]
State
Enables or disables the internal generator. The generator signal is output at the GEN
Output
50
Ω
connector on the front panel.
Remote command:
Level
Defines the output power of the internal generator.
The default output power is -20 dBm. The range is from -60
dBm to +10
dBm.
Remote command:
SOURce<si>:POWer[:LEVel][:IMMediate][:AMPLitude]
Level Offset
Defines an offset to the output power of the internal generator. Used to adapt the level
display, for example to cable loss.
Remote command:
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet
Frequency
Defines the frequency of the internal generator signal as an independent CW source.
The step size depends on the measurement mode.
If the internal generator is used as a tracking generator, the frequency is coupled to the
frequency of the analyzer. Thus, this setting is not available.
Remote command:
Receiving Data Input and Providing Data Output