Remote Commands
R&S
®
FPL1000
756
User Manual 1178.3370.02 ─ 09
Suffix:
<n>
.
Setting parameters:
<WindowType>
Type of measurement window you want to replace another one
with.
See
on page 750 for a list of availa-
ble window types.
Example:
LAY:WIND2:REPL MTAB
Replaces the result display in window 2 with a marker table.
Usage:
Setting only
11.7.2
Examples: Configuring the Result Display
The following example demonstrates how to configure result displays in a remote envi-
ronment.
11.7.2.1
Example 1: Adding and Arranging Windows
Starting from the default initial display in the Spectrum application (Frequency Sweep),
we will configure the following result displays:
1 Frequency Sweep
3 Marker Table
2 Spectrogram
4 Marker Peak List
//--------------Resetting the instrument -----------
*RST
//--------------- Adding new windows --------------------
//Add a Spectrogram window beneath the Frequency Sweep window
LAY:ADD? '1',BEL,SGR
//Result: window number: '2'
//Add a Marker Table window to the right of the Frequency Sweep window
LAY:ADD? '1',RIGH,MTAB
//Result: window number: '3'
Remote Commands for Result Displays