Remote Commands
R&S
®
FPL1000
766
User Manual 1178.3370.02 ─ 09
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:UPP 2
Example:
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFigure:TRIGger
<State>
Defines the behavior of the measurement when adjusting a setting automatically (using
SENS:ADJ:LEV ON
, for example).
See
"Adjusting settings automatically during triggered measurements"
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
ADJ:FREQ
Manual operation:
See
"Adjusting the Center Frequency Automatically (Auto Fre-
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S
FPL1000 or limiting the dynamic range by
an S/N ratio that is too small.
Example:
ADJ:LEV
Manual operation:
See
"Setting the Reference Level Automatically (Auto Level)"
Setting Basic Spectrum RF Measurement Parameters