Remote Commands
R&S
®
FPL1000
660
User Manual 1178.3370.02 ─ 09
<sb>
irrelevant
Example:
CALC:LIM:ESP:REST
Resets the limit lines for the current Spectrum Emission Mask
standard to the default setting.
[SENSe:]ESPectrum<sb>:PRESet[:STANdard]
<Standard>
This command loads a measurement configuration.
Standard definitions are stored in an xml file. The default directory for SEM standards
is
C:\Users\Public\Documents\Rohde-Schwarz\Analyzer\sem_std
.
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
Parameters:
<Standard>
String containing the file name.
If you have stored the file in a subdirectory of the directory men-
tioned above, you have to include the relative path to the file.
Manual operation:
See
See
[SENSe:]ESPectrum<sb>:PRESet:RESTore
This command restores the default configurations of predefined SEM standards.
Note that the command will overwrite customized standards that have the same name
as predefined standards.
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
Manual operation:
See
[SENSe:]ESPectrum<sb>:PRESet:STORe
<Standard>
This command saves the current SEM measurement configuration.
Standard definitions are stored in an xml file. The default directory for SEM standards
is
C:\Users\Public\Documents\Rohde-Schwarz\Analyzer\sem_std
.
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
Parameters:
<Standard>
String containing the file name.
You can save the file in a subdirectory of the directory men-
tioned above. In that case, you have to include the relative path
to the file.
Configuring and Performing Measurements