![Ronde & Schwarz 1304.0004.03 User Manual Download Page 114](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985114.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
114
User Manual 1178.3370.02 ─ 09
The measurement is based on the example
"Measuring the Power of a GSM Burst
1. Switch off the power measurement.
a) Press the [MEAS] key.
b) Select "Zero Span".
2. Increase the time resolution by setting the sweep time to
100
µs
([SWEEP] >
"Sweep Time Manual").
3. Shift the rising edge of the GSM burst to the center of the display by defining a trig-
ger offset.
a) Press the [TRIG] key.
b) Select "Trigger Offset".
c) By turning the rotary knob counterclockwise, reduce the trigger offset until the
burst edge is displayed in the center of the display, or enter
-50
µs
.
The R&S
FPL1000 displays the rising edge of the GSM burst.
Figure 7-16: Rising edge of the GSM burst displayed with high time resolution
4. Move the falling edge of the burst to the center of the display. To do so, switch the
trigger "Slope" to "Falling" ([TRIG] > "Trigger/ Gate Config").
The R&S
FPL1000 displays the falling edge of the GSM burst.
Measurements and Results