Remote Commands
R&S
®
FPL1000
658
User Manual 1178.3370.02 ─ 09
Retrieving results:
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
Remote commands exclusive to occupied bandwidth measurements:
................................................................................................ 658
...........................................................................................658
[SENSe:]POWer:BWIDth
<Percentage>
[SENSe:]POWer:BANDwidth
<Percentage>
This command selects the percentage of the total power that defines the occupied
bandwidth.
Parameters:
<Percentage>
Range:
10 PCT to 99.9 PCT
*RST:
99 PCT
Default unit: PCT
Example:
POW:BAND 95PCT
Manual operation:
See
11.6.6.2
Programming Example: OBW Measurement
This programming example demonstrates the measurement example described in
Chapter 7.2.6.5, "Measurement Example"
on page 159 in a remote environment.
//-----------Preparing the measurement ------------
//Reset the instrument
*RST
//-------------Configuring the Measurement-----------------------
//Set the center frequency to 800 MHz.
FREQ:CENT 800MHz
//Set the reference level to -10 dBm.
DISP:TRAC:Y:RLEV -10dBm
//Activate occupied bandwidth measurement.
CALC:MARK:FUNC:POW:SEL OBW
//Set the percentage of power to 99%.
POW:BWID 99PCT
//Set the channel bandwidth to 21 kHz.
POW:ACH:BAND 21kHz
//Optimize the instrument settings according to the channel bandwidth.
POW:ACH:PRES OBW
//Determine the ideal reference level for the measurement.
POW:ACH:PRES:RLEV
//Set the trace detector to positive peak.
DET APE
//--------------Performing the Measurement-----
//Select single sweep mode.INIT:CONT OFF
Configuring and Performing Measurements