Remote Commands
R&S
®
FPL1000
745
User Manual 1178.3370.02 ─ 09
The R&S
FPL1000 caches all measurement parameters that you can set with this
command. If you use the command repeatedly, the R&S
FPL1000 only changes those
settings that you have actually changed before initiating the measurement. Thus, mea-
surement times are kept as low as possible.
If you synchronize the measurement with
*OPC
, the R&S
FPL1000 produces a service
request when all frequencies have been measured and the number of individual mea-
surements has been performed.
Note that using the command as a query initiates the measurement and returns the
results if all frequencies have been measured. For more information on querying the
results see
Parameters:
<Frequency>
Defines the pulse frequency.
Range:
0 to Fmax
Default unit: Hz
<RBW>
Defines the resolution bandwidth.
Default unit: HZ
<MeasTime>
Defines the measurement time.
Range:
1 µs to 30 s
Default unit: S
<TriggerSource>
Selects a trigger source.
<TriggerLevel>
Defines a trigger level.
The trigger level is available for the video trigger. In that case,
the level is a percentage of the diagram height.
In case of an external trigger, the R&S
FPL1000 uses a fix TTL
level.
Range:
0 to 100
Default unit: PCT
<TriggerOffset>
Defines the trigger delay.
Range:
0 s to 30 s
*RST:
0 s
Default unit: s
<Detector>
Selects the detector and therefore the way the measurement is
evaluated.
MEAN
Calculates the RMS pulse power.
PEAK
Calculates the peak pulse power.
<OfPulses>
Defines the number of pulses included in the measurement.
Range:
1 to 32001
Configuring and Performing Measurements