![Ronde & Schwarz 1304.0004.03 User Manual Download Page 825](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985825.webp)
Remote Commands
R&S
®
FPL1000
825
User Manual 1178.3370.02 ─ 09
CALCulate<n>:SGRam[:STATe]
<State>
CALCulate<n>:SPECtrogram[:STATe]
<State>
This command turns the spectrogram on and off.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:SGR ON
Activates the Spectrogram result display.
CALCulate<n>:SGRam:THReedim[:STATe]
<State>
CALCulate<n>:SPECtrogram:THReedim[:STATe]
<State>
Activates or deactivates a 3-dimensional spectrogram for the selected result display.
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
CALC:SPEC:THR:STAT ON
Manual operation:
See
CALCulate<n>:SGRam:TRACe
<Trace>
CALCulate<n>:SPECtrogram:TRACe
<Trace>
This command determines the trace in the result display the Spectrogram is based on.
Suffix:
<n>
.
Parameters:
<Trace>
TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6
How many traces are available depends on the selected result
display.
Example:
CALC2:SPEC:TRAC TRACE3
Setting Basic Spectrum RF Measurement Parameters