Remote Commands
R&S
®
FPL1000
853
User Manual 1178.3370.02 ─ 09
Suffix:
<n>
.
<m>
Manual operation:
See
CALCulate<n>:MARKer<m>:MINimum:AUTO
<State>
This command turns an automatic marker peak search for a trace minimum on and off.
The R&S
FPL1000 performs the peak search after each sweep.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:MARK:MIN:AUTO ON
Activates the automatic minimum value search function for
marker 1 at the end of each particular sweep.
Manual operation:
See
"Auto Max Peak Search / Auto Min Peak Search"
CALCulate<n>:MARKer<m>:MINimum:LEFT
This command moves a marker to the next minimum value.
The search includes only measurement values to the right of the current marker posi-
tion.
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
Suffix:
<n>
.
<m>
Manual operation:
See
CALCulate<n>:MARKer<m>:MINimum:NEXT
This command moves a marker to the next minimum value.
In the spectrogram, the command moves a marker horizontally to the minimum level in
the currently selected frame. The vertical marker position remains the same.
Setting Basic Spectrum RF Measurement Parameters