Remote Commands
R&S
®
FPL1000
662
User Manual 1178.3370.02 ─ 09
Remote commands exclusive to multi-SEM measurements
..................................................................................662
...................................................................................662
[SENSe:]ESPectrum<sb>:SCENter
<Frequency>
This command defines the center frequency of the selected sub block in a Multi-SEM
measurement.
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
Parameters:
<Frequency>
Frequency within the currently defined global span (see
Range:
1 to 3
*RST:
1
Default unit: Hz
Example:
ESP1:SCEN 1GHZ
Manual operation:
See
[SENSe:]ESPectrum<sb>:SCOunt
<Subblocks>
This command defines the number of sub blocks in the SEM measurement.
Suffix:
<sb>
.
irrelevant
Parameters:
<Subblocks>
Number of sub blocks in the SEM measurement.
Range:
1 to 3
*RST:
1
Example:
ESP:SCO 2
Manual operation:
See
11.6.7.4
Configuring a Sweep List
The following commands define a sweep list for SEM measurements.
The sweep list cannot be configured using remote commands during an on-going
sweep operation.
Configuring and Performing Measurements