Remote Commands
R&S
®
FPL1000
787
User Manual 1178.3370.02 ─ 09
Example:
SWE:EGAT ON
Switches on the gate mode.
SWE:EGAT:TYPE EDGE
Switches on the edge-triggered mode.
SWE:EGAT:HOLD 100US
Sets the gate delay to 100 µs.
SWE:EGAT:LEN 500US
Sets the gate opening time to 500 µs.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
[SENSe:]SWEep:EGATe:HOLDoff
<DelayTime>
This command defines the delay time between the gate signal and the continuation of
the measurement.
Note:
If you perform gated measurements in combination with the IF Power trigger, the
R&S
FPL1000 ignores the holding time for frequency sweep, FFT sweep, zero span
and I/Q mode measurements.
Parameters:
<DelayTime>
Range:
0 s to 30 s
*RST:
0 s
Default unit: S
Example:
SWE:EGAT:HOLD 100us
Manual operation:
See
[SENSe:]SWEep:EGATe:LENGth
<GateLength>
This command defines the gate length.
Parameters:
<GateLength>
Range:
125 ns to 30 s
*RST:
400μs
Default unit: S
Example:
SWE:EGAT:LENG 10ms
Manual operation:
See
[SENSe:]SWEep:EGATe:POLarity
<Polarity>
This command selects the polarity of an external gate signal.
The setting applies both to the edge of an edge-triggered signal and the level of a
level-triggered signal.
Parameters:
<Polarity>
POSitive | NEGative
*RST:
POSitive
Setting Basic Spectrum RF Measurement Parameters