Remote Commands
R&S
®
FPL1000
803
User Manual 1178.3370.02 ─ 09
The command is available after you have created a reference trace for the selected
measurement type with
[SENSe:]CORRection:COLLect[:ACQuire]
This command is only available if the Tracking Generator is active (see
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
1
Example:
CORR ON
Activates normalization.
Manual operation:
See
[SENSe:]CORRection:COLLect[:ACQuire]
<MeasType>
This command initiates a reference measurement (calibration). The reference mea-
surement is the basis for the measurement normalization. The result depends on
whether a reflection measurement or transmission measurement is performed (see
To obtain a correct reference measurement, a complete sweep with synchronization to
the end of the sweep must have been carried out. This is only possible in the single
sweep mode.
This command is only available if the Tracking Generator is active (see
Setting parameters:
<MeasType>
THRough | OPEN
THRough
"TRANsmission" mode: calibration with direct connection
between generator and device input
"REFLection" mode: calibration with short circuit at the input
OPEN
only allowed in "REFLection" mode: calibration with open input
Example:
INIT:CONT OFF
Selects single sweep operation
CORR:METH TRAN
Selects a transmission measurement.
CORR:COLL THR;*WAI
Starts the measurement of reference data using direct connec-
tion between generator and device input and waits for the sweep
end.
Usage:
Setting only
Setting Basic Spectrum RF Measurement Parameters