![Ronde & Schwarz 1304.0004.03 User Manual Download Page 337](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985337.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
337
User Manual 1178.3370.02 ─ 09
External Trigger 1 ← Trigger Source
Data acquisition starts when the TTL signal fed into the trigger input connector of the
R&S
FPL1000 meets or exceeds the specified trigger level.
(See
Remote command:
TRIG:SOUR EXT
See
SWE:EGAT:SOUR EXT
for gated triggering, see
Video ← Trigger Source
Defines triggering by the video signal, i.e. the filtered and detected version of the input
signal (the envelope of the IF signal), as displayed on the screen.
Define a trigger level from 0 % to 100 % of the diagram height. The absolute trigger
level is indicated by a horizontal trigger line in the diagram, which you can also move
graphically to change the trigger level.
A fixed hysteresis of 10
% of the specified trigger value (in V) is applied to the video
trigger level automatically and cannot be changed.
Video mode is only available in the time domain, and not for I/Q-based data.
Remote command:
TRIG:SOUR VID
, see
IF Power ← Trigger Source
The R&S
FPL1000 starts capturing data as soon as the trigger level is exceeded
around the third intermediate frequency.
For frequency sweeps, the third IF represents the start frequency. The trigger band-
width at the third IF depends on the RBW and sweep type.
For measurements on a fixed frequency (e.g. zero span or I/Q measurements), the
third IF represents the center frequency.
This trigger source is only available for RF input.
For details on available trigger levels and trigger bandwidths, see the data sheet.
Note:
Be aware that in auto sweep type mode, due to a possible change in sweep
types, the trigger bandwidth can vary considerably for the same RBW setting.
Remote command:
TRIG:SOUR IFP
, see
SWE:EGAT:SOUR IFP
for gated triggering, see
Time ← Trigger Source
Remote command:
TRIG:SOUR TIME
, see
Trigger Level
Defines the trigger level for the specified trigger source.
Trigger and Gate Configuration