![Ronde & Schwarz 1304.0004.03 User Manual Download Page 617](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985617.webp)
Remote Commands
R&S
®
FPL1000
617
User Manual 1178.3370.02 ─ 09
*TRG
Trigger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal. This common command complements the commands of the
TRIGger
subsystem.
*TRG
corresponds to the
INITiate:IMMediate
Usage:
Event
*TST?
Self-test query
Initiates self-tests of the instrument and returns an error code.
Note:
If you start a self-test remotely, then select the "Local" softkey while the test is
still running, the instrument only returns to the manual operation state after the test is
completed. In this case, the self-test cannot be aborted.
Return values:
<ErrorCode>
integer > 0 (in decimal format)
An error occurred.
0
No errors occurred.
Usage:
Query only
*WAI
Wait to continue
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
).
Usage:
Event
11.4
Commands for Remote Instrument Operation
The following commands are required to shutdown or reboot the R&S
FPL1000 from a
remote PC.
Commands for Remote Instrument Operation