The Spectrum Application (RF Measurements)
R&S
®
FPL1000
267
User Manual 1178.3370.02 ─ 09
Automatic Peak Search
If activated, a peak search is performed automatically for all active markers after each
sweep.
If Auto peak search
and
limit lines are active, the active markers are set to the peak
delta
values between the measured signal and the limit lines.
Note:
Auto Max Peak Search / Auto Min Peak Search
are not available for EMI measurements.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:PEAKsearch:AUTO
Dwell Time
Sets the dwell time for the EMI marker measurement.
For more information see
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DWELl
Final Test Detector
Defines the detector to be used for the final EMI test at the marker frequency.
This setting is only available if the EMI (R&S
FPL1-K54) measurement option is instal-
led.
For details, see
Note:
The trace detector configured in the trace settings is used for the initial peak
search only, see
Chapter 7.10.1.2, "Trace Settings"
"Off"
No final test is performed.
"PositivePeak"
Determines the maximum signal level that was detected during the
specified dwell time.
"Average"
Determines the average signal level of the samples that were collec-
ted during the specified dwell time.
"Quasi-Peak"
Determines the maximum signal level weighted to CISPR
16-1-1 that
was detected during the dwell time.
The "Quasi-Peak" detector is only available for the CISPR filter, and
not for an RBW of 1
MHz.
"CISPR Aver-
age"
Determines a weighted average signal level according to CISPR
16-1-1.
The average value according to CISPR 16-1-1 is the maximum value
detected while calculating the linear average value during the speci-
fied dwell time.
The "CISPR Average" detector is only available for the CISPR filter.
"RMS Aver-
age"
A combination of the RMS detector (for pulse repetition frequencies
above a corner frequency) and the Average detector (for pulse repeti-
tion frequencies below the corner frequency).
Lowpass filters of the second order determine the average value
(simulation of a mechanical pointer instrument).
The "RMS Average" detector is only available for the CISPR filter.
Measurements and Results