![Ronde & Schwarz 1304.0004.03 User Manual Download Page 809](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985809.webp)
Remote Commands
R&S
®
FPL1000
809
User Manual 1178.3370.02 ─ 09
*RST:
IF
Example:
OUTP:IF VID
Selects the video signal for the IF/VIDEO output connector.
Manual operation:
See
OUTPut<up>:IF:IFFRequency
<Frequency>
This command defines the frequency for the IF output of the R&S
FPL1000. The IF fre-
quency of the signal is converted accordingly.
This command is available in the time domain and if the IF/VIDEO output is configured
for IF.
Suffix:
<up>
.
Parameters:
<Frequency>
*RST:
50.0 MHz
Default unit: HZ
Manual operation:
See
OUTPut<up>:UPORt:STATe
<State>
This command toggles the control lines of the user ports for the
AUX PORT
connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S
FPL1000.
Suffix:
<up>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
User port is switched to INPut
ON | 1
User port is switched to OUTPut
Example:
OUTP:UPOR:STAT ON
OUTPut<up>:UPORt[:VALue]
<Value>
This command sets the control lines of the user ports.
The assignment of the pin numbers to the bits is as follows:
Bit
7
6
5
4
3
2
1
0
Pin
N/A
N/A
5
3
4
7
6
2
Bits 7 and 6 are not assigned to pins and must always be 0.
The user port is written to with the given binary pattern.
Setting Basic Spectrum RF Measurement Parameters