Remote Commands
R&S
®
FPL1000
806
User Manual 1178.3370.02 ─ 09
FIXed
Fixed power level. Required for tracking generator.
SWEep
Increasing power level. Required for power sweep.
*RST:
FIXed
Example:
SOUR:POW:MODE:SWE
SOUR:POW:STAR -50
SOUR:POW:STOP -10
Manual operation:
See
See
SOURce<si>:POWer:STARt
<StartLevel>
Defines the start power level of the internal generator signal for power sweeps. This
value must be lower than the stop level (see
Suffix:
<si>
.
irrelevant
Parameters:
<StartLevel>
Range:
-60 to +10
*RST:
-30
Default unit: DBM
Example:
SOUR:POW:MODE:SWE
SOUR:POW:STAR -50
SOUR:POW:STOP -10
Manual operation:
See
SOURce<si>:POWer:STOP
<StopLevel>
Defines the end power level of the internal generator signal for power sweeps. This
value must be higher than the start level (see
Suffix:
<si>
.
irrelevant
Parameters:
<StopLevel>
Range:
-60 to +10
*RST:
-20
Default unit: DBM
Example:
SOUR:POW:MODE:SWE
SOUR:POW:STAR -50
SOUR:POW:STOP -10
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters