![Ronde & Schwarz 1304.0004.03 User Manual Download Page 338](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985338.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
338
User Manual 1178.3370.02 ─ 09
For gated measurements, this setting also defines the gate level.
For details on supported trigger levels, see the data sheet.
Remote command:
TRIGger[:SEQuence]:LEVel:IFPower
TRIGger[:SEQuence]:LEVel:IQPower
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]
Repetition Interval
Defines the repetition interval for a time trigger. The shortest interval is 2 ms.
The repetition interval should be set to the exact pulse period, burst length, frame
length or other repetitive signal characteristic.
Remote command:
TRIGger[:SEQuence]:TIME:RINTerval
Drop-Out Time
Defines the time the input signal must stay below the trigger level before triggering
again.
For more information on the drop-out time, see
Remote command:
Trigger Offset
Defines the time offset between the trigger event and the start of the sweep.
For more information, see
Offset > 0:
Start of the sweep is delayed
Offset < 0:
Sweep starts earlier (pretrigger)
Only possible for zero span (e.g. I/Q Analyzer application) and gated trigger switched off
Maximum allowed range limited by the sweep time:
Pretrigger
max
= sweep time
max
For the "Time" trigger source, this function is not available.
Remote command:
TRIGger[:SEQuence]:HOLDoff[:TIME]
Hysteresis
Defines the distance in dB to the trigger level that the trigger source must exceed
before a trigger event occurs. Setting a hysteresis avoids unwanted trigger events
caused by noise oscillation around the trigger level.
This setting is only available for "IF Power" trigger sources. The range of the value is
between 3
dB and 50
dB with a step width of 1
dB.
For more information, see
Remote command:
TRIGger[:SEQuence]:IFPower:HYSTeresis
Trigger and Gate Configuration