![Ronde & Schwarz 1304.0004.03 User Manual Download Page 647](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985647.webp)
Remote Commands
R&S
®
FPL1000
647
User Manual 1178.3370.02 ─ 09
Parameters:
<ChannelNumber>
Range:
1 to 18
*RST:
1
Manual operation:
See
11.6.4.5
Checking Limits
The following commands configure and query limit checks for channel power and
ACLR measurements.
The results of the power limit checks are also indicated in the
STAT:QUES:ACPL
sta-
tus registry (see
"STATus:QUEStionable:ACPLimit Register"
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute
..................................................... 647
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute:STATe
........................................... 648
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]
....................................................648
CALCulate<n>:LIMit<li>:ACPower:ACHannel:RESult?
.......................................................648
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]:STATe
..........................................649
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute
.............................................. 649
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute:STATe
.................................... 650
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]
.............................................650
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:RESult?
................................................651
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]:STATe
...................................652
CALCulate<n>:LIMit<li>:ACPower[:STATe]
....................................................................... 652
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute
<LowerLimit>[,
<UpperLimit>]
This command defines the absolute limit of the adjacent channels.
If you have defined an absolute limit as well as a relative limit, the R&S
FPL1000 uses
the lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Parameters:
<LowerLimit>
The limit of the lower adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Default unit: dBm
<UpperLimit>
The limit of the upper adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Default unit: dBm
Manual operation:
See
Configuring and Performing Measurements