![Ronde & Schwarz 1304.0004.03 User Manual Download Page 717](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985717.webp)
Remote Commands
R&S
®
FPL1000
717
User Manual 1178.3370.02 ─ 09
Measuring the RMS Power
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:AVERage:RESult?
This command queries the average RMS of the time domain power. The query is only
possible if averaging has been activated previously using
MARKer<m>:FUNCtion:SUMMary:AVERage
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<RMSPower>
RMS power of the signal during the measurement time.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:PHOLd:RESult?
This command queries the maximum RMS of the time domain power. The query is only
possible if the peak hold function has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<RMSPower>
RMS power of the signal during the measurement time.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:RESult?
This command queries the RMS of the time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
Configuring and Performing Measurements