Remote Commands
R&S
®
FPL1000
764
User Manual 1178.3370.02 ─ 09
[SENSe:]ADJust:CONFigure:DURation
............................................................................ 764
[SENSe:]ADJust:CONFigure:DURation:MODE
..................................................................765
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
.............................................................. 765
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
............................................................... 765
[SENSe:]ADJust:CONFigure:TRIGger
..............................................................................766
..........................................................................................766
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Center frequency
●
Reference level
Example:
ADJ:ALL
Manual operation:
See
"Adjusting all Determinable Settings Automatically (Auto
[SENSe:]ADJust:CONFigure:DURation
<Duration>
In order to determine the ideal reference level, the R&S
FPL1000 performs a measure-
ment on the current input data. This command defines the length of the measurement
if
[SENSe:]ADJust:CONFigure:DURation:MODE
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation:
See
"Changing the Automatic Measurement Time (Meastime
Setting Basic Spectrum RF Measurement Parameters