The Spectrum Application (RF Measurements)
R&S
®
FPL1000
185
User Manual 1178.3370.02 ─ 09
File names must be compatible with the Windows conventions for file names. In partic-
ular, they must not contain special characters such as ":", "*", "?".
For details on the filename and location, see
Chapter 8.2.2.2, "Storage Location and
Load Standard
Loads the selected measurement settings file.
Remote command:
[SENSe:]ESPectrum<sb>:PRESet[:STANdard]
File Explorer
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
Save Standard
Saves the current measurement settings for a specific standard as a file with the
defined name.
Remote command:
[SENSe:]ESPectrum<sb>:PRESet:STORe
Delete Standard
Deletes the selected standard. Standards predefined by Rohde & Schwarz can also be
deleted. A confirmation query is displayed to avoid unintentional deletion of the stan-
dard.
Note:
Restoring predefined standard files. The standards predefined by Rohde &
Schwarz available at the time of delivery can be restored using the "Restore Standard
Files" function (see
Restore Standard Files
Restores the standards predefined by Rohde & Schwarz available at the time of deliv-
ery.
The XML files from the
C:\ProgramData\Rohde-Schwarz\ZNL-FPL\sem_backup
folder are copied to the
C:\ProgramData\Rohde-Schwarz\ZNL-FPL\sem_std
folder.
Note that this function will overwrite customized standards that have the same name
as predefined standards.
Remote command:
[SENSe:]ESPectrum<sb>:PRESet:RESTore
List Evaluation (Results Configuration)
Access
: "Overview" > "Select Measurement" > "Spectrum Emission Mask" > "List
Evaluation"
In the "List Evaluation" dialog box, you configure the contents and display of the SEM
results.