![Ronde & Schwarz 1304.0004.03 User Manual Download Page 724](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985724.webp)
Remote Commands
R&S
®
FPL1000
724
User Manual 1178.3370.02 ─ 09
INIT;*WAI
//Initiates the measurement and finishes the sweep.
//-------------Retrieving the Results------------------------
CALC:MARK:FUNC:HARM:LIST?
//Queries the position of the harmonics.
CALC:MARK:FUNC:HARM:DIST? TOT
//Queries the total harmonic distortion.
11.6.12
Measuring the Third Order Intercept Point
..............................................................................................724
Programming Example: Measuring the TOI
..........................................................727
11.6.12.1
Determining the TOI
All remote control commands specific to TOI measurements are described here.
Useful commands for TOI measurements described elsewhere
●
●
CALCulate<n>:DELTamarker<m>:X:RELative?
●
●
●
Remote commands exclusive to TOI measurements
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe]
.......................................................... 724
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal ONCE
...................................... 725
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult?
......................................................... 725
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult:MAXimum?
......................................... 725
CALCulate<n>:MARKer<m>:FUNCtion:TOI:RESult:MINimum?
.......................................... 726
CALCulate<n>:MARKer<m>:FUNCtion:TOI[:STATe]
<State>
This command initiates a measurement to determine the third intercept point.
A two-tone signal with equal carrier levels is expected at the RF input of the instru-
ment. Marker 1 and marker 2 (both normal markers) are set to the maximum of the two
signals. Delta marker 3 and delta marker 4 are positioned to the intermodulation prod-
ucts. The delta markers can be modified separately afterwards with the
command.
The third-order intercept is calculated from the level spacing between the normal mark-
ers and the delta markers.
Suffix:
<n>
.
Configuring and Performing Measurements